Cart (0 item items)

RCP: Reducing Compound Photometer

For the Detection of Hydrogen, Carbon Monoxide & Select Hydrocarbons

Please contact us directly to purchase this item

Simple, accurate measurements.

The Peak Performer 1 uses a reducing compound photometer (RCP) to measure trace impurities down to part per billion levels, while offering a wide linear range for simple, accurate measurements.

The PP1 technology has proven to offer a more cost-effective and user friendly alternative to other instruments for similar applications. In addition, Peak Labs practical experience and ability to customize to suit your applications makes Peak Labs your partner, not just your supplier.

Technical Information

The Peak Performer 1 (PP1) gas chromatograph (GC) can be optimized for your analytical needs in a variety of matrix gases.

Accuracy

  • + / - detection limit or 10% of reading, whichever is higher.

Range

  • 1000 : 1
    Examples
    1 ppb – 1 ppm w/1m L sample loop
    10 ppb – 10 ppm w/100 ul sample loop

Operation

Run time:

  • 2 < minutes (depending on application)

Operating Temperature:

  • 55 - 85 °F (13 - 30 °C )

Carrier Gas Supply:

  • 99.999% or purified grade to < 1 ppm total impurities
  • Supply pressure 70 - 110 psig with 5% stability

Data Collection / Communication (Standard)

  • Backlit 3" x 5.25" touch screen
  • Visual chromatogram and numerical results
  • On-board integration with re-run capability
  • Chromatograms collected and archived via standard 9 – pin com-port in .csv format (data is excel compatible)

Dimensions / Electrical

  • 27" L x 17" W x 7" H
  • 25 lbs.
  • 115 VAC, 50 – 60 Hz / 220 VAC, 50-60 Hz
  • 1.5 amp maximum

Accessories and other features

  • Trend outputs
  • Analysis averaging
  • Auto start/stop capability
  • On column syringe injector adapter
  • Dual sample stream option
  • 4-20 mA Output
Some of our customers